Accéder au contenu:

Résumé de recherche

Objective

This study compares an Impulse Oscillometry System (IOS; CareFusion) and an Airwave Oscillometry System (tremoFlo; THORASYS) in phantoms and humans.

Methods

We built and characterized 3 phantoms representative of normal and obstructed human respiratory systems. Each was measured thrice per device, and averages of resistance (R) and reactance (X) versus frequency (IOS: 5-35 Hz; tremoFlo: 5-37 Hz) were calculated. We computed derived parameters and maximum deviations from predicted values (εmax). Six volunteers (4 normal, 2 asthmatic) were also measured per device.

Results

The IOS overestimated predicted R of phantoms up to 36% (Table), or 26% after supplemental filter compensation (FC), and the area under the reactance curve (AX) by 31%. For tremoFlo, εmax peaked at 5.3%. In subjects, IOS resistance at 5 Hz (R5) was higher than tremoFlo pre-FC (21% p=0.066). No significant differences were observed post-FC.

Discussion

IOS results may be affected by anti-bacterial filters. Once this was compensated, we observed no significant differences between IOS and tremoFlo in our small group of normal and mildly abnormal subjects. However, measurements of higher impedance phantoms suggest that the IOS may over-estimate R, AX and Fres in presence of more severe obstruction.


Mots clés
Lung function testing, lung mechanics
Source

Drapeau G, Schuessler T. Oscillometry system comparison in phantoms and humans. Eur Respir J. 2015;46(suppl 59)

Nous contacter
En savoir plus sur la façon dont THORASYS peut vous aider dans votre cabinet. Remplissez le formulaire et l'un de nos représentants prendra contact avec vous.
Formulaire soumis avec succès !
Nous vous contacterons sous peu.